The time range displayed on the JUMO LOGOSCREEN 500 cf is determined by the programmed memory cycle.
Different memory cycles can be programmed for the three possible operating modes: "Normal operation", "Time operation" and "Event operation". This allows, for instance, to save measured values more frequently, e.g. in the event of overrange or in a specific time frame, and, thus, have available a higher resolution within the time axis.
Do the interface parameters (log, baud rate, data format, device address, etc.) in the communication server program JUMO PCC correspond with the settings in the JUMO LOGOSCREEN 500cf?
If necessary, refer to the operating manual, chapter 4.2.10 (Configuration - Interface).
Is the connection, interface type (RS232/RS485) and connection, correct?
For the RS485, an additional external interface converter to RS232 or USB is often required for connection to a PC.
The recording time depends on several factors:
Please refer to chapter 2.8 in this operating manual for the orientation values listed in a table.
Take into consideration, that the internal memory in the Logoscreen 500cf operates as a ring memory, whereas the measured data saved on the external CF card cannot be overwritten automatically by the Logoscreen 500cf.
Thus, the old data remains saved on the CF card if the recording time is exceeded, whereas the internal memory contains the latest data. Measured data is lost for the time inbetween.
Compact Flash memory cards (Industrial Grade) up to max. 2 GB can be used.
Format: FAT or FAT16
For a high data security, we recommend to only use the CF cards available from JUMO as accessories for this recorder. These CF cards were tested together with the Logoscreen 500cf and have been proven in practice.
1. Release of integrators and counters as well as the maths and logics module.
Release of these software options at a later time is possible with the setup program and also directly on the device using the keys.
1.1 Proceed as follows to release the options using the keys on the device:
A code number is read out first from the logoscreen 500cf. For this purpose, invoke the following menu:
Configuration level — Device data — Option release — Determined code No.
The number code displayed here including the possible prefix is sent to JUMO by telephone, fax or mail.
JUMO will generate a device-specific release key from this code No.
Then, enter this release key into the respective field on the device to release the option. The option is immediately available and can be configured and used as required by the user.
1.2 Option release via the setup program:
The extra code "Maths", "Counters/integrators" can be released at a later time using the Logoscreen 500cf setup program and an interface connection to the recorder.
Complete wiring of the Logoscreen 500cf is not required for the option release. It is only necessary to connect the recorder voltage supply.
The interface connection can be easily established with a JUMO USB-TTL or JUMO RS232-TTL interface via the serial programming interface.
An optional RS485/RS422/RS232 interface can be used for establishing the connection as an alternative to the JUMO interface lines.
A code number for maths and counters/integrators can be generated (= read out) in the setup program under the "EXTRAS" - "Release of extra codes" menu point. The displayed code number including the possible prefix is sent to JUMO by telephone, fax or mail.
JUMO will generate a device-specific release code from this code number.
Enter this release code in the setup program in the field following "Enter release code" on the PC to release the option.
2. Upgrade from three to six analog inputs
If required, Logoscreen 500cf delivered with three analog inputs can be upgrades at a later date to six analog inputs by the JUMO service personnel.
3. Upgrade of 4 binary inputs, 3 relays and a serial interface RS 232/485 (modbus, J-Bus)
If required, the system can be upgraded at a later time to the extra code "261“ (gr.eq 4 binary inputs, 3 relay outputs, serial interface RS232/485) by the JUMO service personnel.
A malfunction is pending that is precisely displayed in the device info window.
(e.g. "Battery flat“)
Furthermore, general information about the device, e.g. software version number and hardware equipment is displayed in the device info window.
Also refer to the operating manual, chapter 3.7.
An inspection certificate 2.2 of an acceptance test certificate 3.1.B can be issued for the LOGOSCREEN 500cf upon request.
For this purpose, send the device to JUMO including a list of the specified measuring points to be moved to.
It is possible to calibrate complete temperature measuring strings in the user factory as well as in our plant in addition to the calibration of the individual paperless recorder. For this purpose, a DKD laboratory is available in the JUMO main factory in Fulda.
Up to 6 analog signals can be recorded simultaneously by the Logoscreen 500cf. This also applies to devices with released maths option.
With the version equipped with 3 analog inputs and released maths module, for example, it is possible to record 3 values calculated in the maths module in addition to the three analog inputs.
Über die Konfigurationsebene am Gerät und das Untermenü „Feinabgleich“ ist es möglich, jeden einzelnen Analogeingang an Referenzsignale anzupassen. Um zu vermeiden, dass dieser gerätespezifische Feinabgleich unbeabsichtigt überschrieben oder übertragen wird, ist er ausschließlich direkt über die Tasten am Schreiber vorzunehmen.
In der Betriebsanleitung Kapitel 4.2.11 ist die genaue Vorgehensweise beschrieben.
Für alle drei Softwareprodukte gilt:
Die jeweils aktuelle Version steht auf der JUMO-Homepage zum kostenlosen Download bereit. Für die Nutzung als Vollversion wird jeweils eine kostenpflichtige Lizenznummer benötigt. Beim Update bleibt die erworbene Lizenznummer weiterhin gültig.
Die PCA3000 kann ohne Lizenznummer zum Anzeigen von PCA-Archivdaten verwendet werden, wenn Messwertaufzeichnungen z.B. an Kunden oder Behörden weiterzugeben sind.
Über die undokumentierten Parameter sind weitergehende Anpassungen der Gerätefunktion an besondere Anforderungen möglich. Diese Parameter sollten jedoch nur nach Rücksprache mit JUMO und auf ausdrückliche Anweisung geändert werden, um Fehleinstellungen zu vermeiden.
Die Voreinstellung bzw. das Zurücksetzen von Zählern und Integratoren ist über das Menü „Zähler-/Int.-Reset“ in der Parametrierebene direkt am Logoscreen 500 cf möglich.
Dort können einfach neue Vorgaben für die einzelnen Zähler und Integratoren gemacht werden. Der Eingriff wird in der Ereignisliste dokumentiert.
Das Kapitel 2.5.2 der Betriebsanleitung enthält dazu weitere Informationen.
Wenn pro Eingangssignal am Schreiber oder bei der Auswertung in PCA3000 zwei Linien bzw. zwei Spalten erscheinen, hat dies folgenden Grund:
Werkseitig ist unter „Registrierung“ - „Normalbetrieb“ - „Speicher-Werte:“ die Einstellung Spitzenwerte gewählt. Dadurch wird in jedem Speicherzyklus der Min- und der Max-Wert ermittelt und gespeichert.
Diese Spitzenwertregistrierung ist besonders für Applikationen ideal, für die folgende Aussagen zutreffen:
Für viele „Standardapplikationen“ empfehlen wir unter „Speicher-Werte:“ die Einstellung
Mittelwert zu wählen. Der Logoscreen 500 cf berechnet dann jeweils vor der Speicherung den Mittelwert aus den im Speicher-Zyklus aufgelaufenen Messwerten und speichert dann nur einen Wert pro Zyklus. Damit wird dann auch nur noch eine Linie pro Kanal dargestellt.